Customization: | Available |
---|---|
After-sales Service: | Lifetime Maintenance |
Power Supply: | 220V |
Still deciding? Get samples of US$ 999/Piece
Order Sample
|
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() ![]() |
---|---|
Support payments in USD |
Secure payments: | Every payment you make on Made-in-China.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Product model: | HJ-GK-100 | ||||||
Nominal content product: | 150L | ||||||
Inner box size: | W500*H600*D500mm | ||||||
Dimensions (approx.) : | W1000*H1750*D1200mm | ||||||
Weight (approx) : | 600KG | ||||||
Temperature range: | (-55-150°С) | ||||||
-160ºC,150ºC,120ºC, -100ºC, -80ºC,-73ºC,-60ºC, -40ºC, -20ºC, 0ºC | |||||||
"+150ºC.200ºC.250ºC.300ºC.500ºC,600ºC.700ºC.800ºC.900ºC.1000ºC.1200ºC (Optional) | |||||||
Vacuum degree: | (10KPA- normal pressure) | ||||||
Atmospheric pressure~1kPa:~133pa.0pa:~ 0. 1MPa:~760*10-8mmHg:0.05Tom | |||||||
1*10-5Pa,1*10-7mbar (optional), vacuum within 2-10 minutes | |||||||
Case leakage rate: | 1*10-8 mbar L/S | ||||||
Humidity range: | RH20%~98% (optional 10%RH ~98%RH or 5%RH ~98%RH) | ||||||
Control accuracy: | Temperature: ±0.2ºC Humidity: ±2.5% (refers to the difference between the controller set value and the controller measured value) | ||||||
Temperature fluctuation: | ≤0.5ºC (temperature fluctuation is half of the difference between the highest temperature and the lowest temperature measured at the center point) | ||||||
Temperature error: | ≤±1 °C (the average temperature shown by the studio temperature controller minus the average temperature measured at the center point) | ||||||
Temperature uniformity: | ≤5.0 °C (Temperature uniformity is the arithmetic mean of the difference between the highest temperature and the lowest temperature measured in each test) | ||||||
Heating rate: | 5ºC/Min (nonlinear no-load), optional: 3ºC/min; 5C/min:8C/min; 10C/min; | ||||||
15ºC/min:20°ºC/min; 30ºC/min(linear or nonlinear) | |||||||
Cooling rate: | 5°C /Min (nonlinear no-load),optional: 3ºC /min; 5ºC/min:8ºC/min;10ºC/min; | ||||||
15ºC/min; 20°C/min :30C/min (linear or non-linear) | |||||||
Relative humidity error: | ±2.5%RH | ||||||
Working noise: | A sound level 70dB(A) (measured in a soundproof chamber with low echo and a ring temperature of 25ºC; The average value of 8 points is tested using A weight. Each test point is 1 meter above the noise source and 1 meter above the ground. |