Vacuum ovens are widely used in biology, chemistry, chemical pharmaceuticals, medical and health, agricultural research, aviation research, satellite research, teaching research, lithium battery research and other industries, powder drying, product baking, vacuum temperature testing, etc., as well as disinfection and sterilization of various glass containers. They are particularly suitable for fast and efficient drying of heat-sensitive, easily decomposable, easily oxidizable substances and items with complex components.
Inner dimensions | Width 850×Depth 800×Height 350mm; (three layers independently controlled) |
Outer dimensions approx | Width 1700×Depth 1120×Height 2150mm including casters, subject to the actual product. |
Interlayer | The single box is divided into 3 fixed layers, 50mm from the bottom and above the first layer, and equipped with 3 mesh plates, which can be pulled out. |
Inner material | 304# stainless steel industrial plate, 5.0mm thick, reinforced treatment, vacuuming without deformation. |
Outer material | Q235 cold-rolled steel plate, 1.2mm thick standard plate, surface powder paint treatment. |
Insulation material | High temperature resistant glass fiber rock wool, thermal insulation effect Good. |
Sealing material | High temperature resistant silicone sealing strip. |
Box structure | Integral type, the left part is upper, middle and lower product boxes, the right part is for electric control and vacuum pump installation, movable casters and adjustable fixed feet are installed at the bottom of the machine, which is convenient for movement and fixed position. |
Timing function | The touch screen instrument has a built-in timing function, which stops heating, vacuum, nitrogen filling and other functions when the time is up, and alarms at the same time. |
Current controller | Solid state relay SSR has no contact |
Heating method | Thermal radiation convection, heating tube is installed outside the cavity. |
Heating element | High temperature resistant stainless steel heating tube. |
Temperature range | The temperature is adjustable from +30 to 200ºC at normal temperature. |
Instrument control accuracy | ±0.5ºC (Constant state) |
Instrument display accuracy | 0.1ºC |
Temperature deviation | ±5ºC. (Empty box test geometric center test point) |
Vacuum range | 0.0KPa to -100.00KPa, the displayed value is different in different regions; |
Vacuum pressure leakage rate | Average about 0.8KPa/hour |
Vacuum control method | Automatic control, stop vacuuming when reaching the upper limit, automatically replenish vacuum when below the lower limit (start difference 0.5KPa interval), so as to work in a cycle and keep the vacuum degree automatically constant. |
Pressure relief method | Adopt solenoid valve control, manual pressure relief when pressure relief is required. |
Vacuum pipeline | Stainless steel pipe, steel wire hose, clamp, etc. |
Vacuum control valve | KF16 high vacuum pneumatic valve, inflation pressure relief valve, pipeline exhaust valve, etc. |
Vacuum configuration | Adopt RH0040 single-stage vacuum pump, power 380V/2.2KW. |
Voltage | AC 380V/50Hz three-phase five-wire |
Current | 30A |
Total power | 20KW |
Compressed gas source | 0.4~0.6MPa vacuum valve use |
Nitrogen source | 0.2~0.3MPa baking product use |